one) Fill in the use file in time following functioning the glove box, such as "user, water oxygen price, what to do"


two) When changing gasoline cylinders in the glove custom blove box, the pressure is set to the default benefit ± 3mbar


three) The daily operating strain is set in accordance to the functioning comfort and ease. When not in use for a very long time, the stress is set to one-6mbar


four) Consistently look at the put on of the gloves, look at the oil level of the vacuum pump (among the upstream and downstream markings), and add the pump oil appropriately


five) Clear the h2o probe as soon as just about every two to 3 months to ensure precise detection. (Also will need to refer to the real condition of the surroundings in distinct person packing containers


six) If you often use organic and natural solvents in the glove box, it is encouraged that you regularly exchange the organic solvent adsorbent substance, not afterwards than 1 yr at the most up-to-date


7) Swap the filter element on a regular basis, no later than two many years


eight) In accordance to the genuine use of your glove box, determine to change the vacuum pump oil cycle, these as black and yellow pump oil, water or more natural solvents in the pump oil. The oil substitute cycle must not exceed two yrs at the most current


9) If a substantial sum of natural solvent is inevitably exposed during the experiment, the cycle wants to be stopped for the experiment. After the experiment, first clean and then get started the cycle


ten) When the setting in the box is toxic or destructive fuel, the exhaust port of the vacuum pump, the exhaust port of the regeneration gasoline and the exhaust port of the cleaning valve should really be fatigued to the out of doors with an exhaust pipe (when exhaust is not permitted to be outside, the exhaust gasoline procedure must be performed to start with).